Welcome to Forever And a Day Technology

  • Mon-Fri: 8.00am to 5.00pm
Search

Ultrasonic Wall Thickness Gauge | TI-007DLX

Ultrasonic Wall Thickness Gauge | TI-007DLX
The TI-007DLX is a precision ultrasonic wall thickness gauge designed
for high-resolution measurements on thin-wall metal and plastics.
Range with Standard Delay Line Transducer
0.0060″ – 1.000″ / 0.150 – 25.40 mm
Range with Contact Transducers
up to 36.00″ / 915.0 mm

Ultrasonic Wall Thickness Gauge | TI-007DLX

Range with Standard Delay Line Transducer
0.0060″ – 1.000″ / 0.150 – 25.40 mm

Range with Contact Transducers
up to 36.00″ / 915.0 mm

 

   The TI-007DLX is a precision ultrasonic wall thickness gauge designed for high-resolution measurements on thin-wall metal and plastics. It is packaged in a sturdy extruded aluminum housing which is sealed for excellent environmental protection assuring trouble-free use even in the toughest field and production environments.

   The built-in datalogger has a capacity of 10,000 data values stored in up to 40 batches with 250 values/batch which is transferred via the USB port using the supplied cable. The TI-007DLX does not require any software, it will be shown as a drive when connected to a PC allowing convenient drag-n-drop into your program of choice. Data is saved in a comma deliminated file format (.csv) which makes it compatible with all programs and software.

   The TI-007DLX has measuring range in steel from 0.0060 to 1.0000 inch, (0.15 to 25.40 mm). Using a single element delay tip transducer, the gauges will measure thin materials and automatically remove any coatings in Echo-Echo Mode (E-E) as well as automatically switch to Interface Echo Mode (I-E)when measuring thicker materials, For measurement of plastics, the user would select Plastic Mode (PLAS) and use the option Graphite Delay Line Tip (p/n X-633-0000).

   Additional Measuring Modes include Differential (DIFF) where a user sets a “nominal value” and the gauge displays the ± difference between then nominal and actual wall thickness, Alarm (user-set Hi/Low Limits) trigger Hi/Lo LED’s when the measured values are out of tolerance and Velocity Mode

Features

      • Includes NIST Traceable Calibration Certificate
      • Large, backlit LCD display retains last reading
      • Resolution of 0.0001 inch (0.001 mm) with Selectable units of inch or mm
      • Built-in datalogging for 10,000 data values stored in up to 40 batches with 250 values/batch with USB Output
      • Scan mode, 100 readings/sec. Probe is slid over a large measuring area, the minimum thickness during the “scan” will be displayed
      • Adjustable GAIN (AGC) in 5 steps to fine tune the gauge for use on challenging materials or applications
      • Can be used with both Single Element Delay Line and contact transducers (5 to 20MHz)
      • LCD shows Thickness Value, Velocity Setting, Gain Setting, Stability & Battery Indicators, Scan Mode, Zero and units
      • Operates on two (2) AA Batteries – 45 hours continuous operation
      • Calibrate to a sample of known thickness which automatically calculates the optimal acoustic velocity
      • Powered by: 120MHz FPGA timing & 100 volt spike pulser
      • Measure Modes: Pulse-Echo (P-E), Echo-Echo (E-E), Interface Echo (I-E) and Plastics (PLAS)
      • Additional Modes Include Differential (from user-entered “nominal value”), Alarm (user-set Hi/Low Limits) and Velocity Mode
      • Optional serial RS-232 or Bluetooth output
      • IP 65 Rated for protection even in the toughest field conditions
      • CE-Certified

Delay Line Tips for TI-007DLX

P/N Description Standard / Optional Image
F-000-7102 1/4″ x 3/8″ for all metals (Short) Standard
F-000-7103 1/4″ x 1/2″ for all metals (Long) Standard
X-633-0000 1/4″ x 3/8″ Graphite Tip for thin plastics (Graphite Short) OPTIONAL

Specifications

Range with Standard Probe0.0060-1.0000″ (0.15-25.40 mm) with standard delay line transducer in steel
Resolution0.0001″ (0.001 mm) . 0.001″ (0.01 mm), user-selected units and resolution
DisplayMulti-function 7 segment 4.5 digit liquid crystal display with Back-light
Velocity Range0.0120 to .7300 in/μs (305 to 18,542 m/sec)
Display BacklightBacklight is selectable on/off/auto, and selectable brightness (Lo, Med, Hi) options
Output OptionalRS-232 and Bluetooth
Cable4 ft. (1.2 m) with Microdot/Lemo00 connector.
OutputUSB single value only (no internal memory)TI-007DLX for a model with build-in memory for 40,000 values USB output
Operating Temperature-22 to 167F (-30 to 75C)
Housing ProtectionIP65
Battery LifeTypically operates for 35 hours on alkaline and 18 hours on NiCad
Weight11 ounces (308 grams)
Dimensions2.5 x 5.17 x 1.25″
(64 x 131 x 32 mm)
WarrantyGauge: 5 Years
Probes: 90 Days
Total Measuring Range
(Steel)
0.006 – 36.000 inches
0.15 – 924 mm
(depends on material and transducer/probe type)
Measuring ModesPulse-Echo (P-E), Echo-Echo (E-E), Interface Echo (I-E) and Plastics (PLAS), Differential, Alarm, Scan and Velocity (VX)
GainAutomatic time dependent gain (TDG) with manual override
Probe (Standard)1/4″ diameter, 15 MHz delay line transducer with cable (p/n T-402-5507)
Probes (optional)Various Single Element Delay Line and Contact Transducers (5 to 20MHz)
Display Update Rate10 Hz (10 updates/sec)
Battery Type2x AA 1.5V alkaline, 1.2V NiCad, or 1.5V lithium AA cells (rechargeable batteries can be used)
HousingExtruded aluminum body with nickel-plated aluminum end caps (gasket sealed)
CertificationsNIST Traceable and MIL-STD-45662A
KeypadSealed membrane that is resistant to both water and petroleum products
Seven or eight tactile-feedback keys
Pulse Repetition Frequency (PRF)200 Hz (200 pulses/sec)
To Lock SettingsWith gauge powered off, press and hold the “+” key while powering on the gauge
The LOCK icon will be illuminated on the Display
To unLock SettingsWith gauge powered off, press and hold the “-” key while powering on the gauge
The LOCK icon will turn off